Sword rh Probes

Models HC2-HS28 and HC2-HS42

These relative humidity and temperature probes are designed for the paper and printing industry. The moisture content of paper can vary widely and have a dramatic effect on the quality of printing. Careful control of the moisture content of printing paper can improve paper feed and the efficiency of the printing process.

 

SPECIFICATIONS

Range of application

-40 to +85 °C / 0 to 100 %rh

Digital interface (UART)

& scalable analog outputs

0 to 1 V

Standard output scaling

0 to 1 V = -40 to +60 °C / 0 to 100 %rh

Response time

less than 15 s

Accuracy

±0.8 %rh, ±0.1 K, at 23 °C ±5 K

Power supply

3.2 to 5 VDC, calibrated at 3.3 VDC, current: approx. 4.5 mA

Sensor type

ROTRONIC HYGROMER® IN1, Pt100 Class A

Filter type

No filter

Material

Aluminium (probe), POM (handle)

 

HC2-HS28

HC2-HS42

Probe length

280 mm

420 mm

Weight

220 g

240 g

Adjusted at 23 °C and 10, 35, 80 %rh

COMPATIBLE PRODUCTS

Transmitters

HygroFlex HF5, HygroFlex HF8

Handheld instruments

HygroPalm HP22-A, HygroPalm HP23-A

Data loggers

HygroLog HL-NT2, HygroLog HL-NT3,

INCLUDED

Factory calibration certificate

Short instruction manual (BFC-UART)

TYPICAL ACCESSORIES

Replacement filter (BFC-UART)

ET-W37-Set

Calibration device

EGS

Humidity standard for calibration:

                    10 %rh

                    35 %rh

                    80 %rh

 

 

EA10-SCS

EA35-SCS

EA80-SCS

relative humidity and temperature sensor HC2-HS28 and HC2-HS42