SpectraSensors Laser Hygrometer - Integrated System
Sample System Element
|
Pressure Letdown |
Upto 3 stage heated regulators |
|
Pressure Indication |
Standard Bourdon Gauges, option pressure transmitters |
|
Filtration |
Particulate |
|
First Stage Coalescor |
|
Second Stage Membrane Type |
|
Flow Control |
Needle Valve |
|
Flow Measurement |
Armoured Flow Meter, optional flow alarms |
|
Analyser Protection |
Pressure Relief after each regulator stage |
|
|
Excess Flow Valve Before Analyser |
Mechanical
|
Material |
GRP with integral insulation and Stainless Steel
Fittings |
|
IP Rating |
IP66 |
|
Ambient |
-40 to +40 oC |
|
Heaters |
Upto 1200 Watts, suitable for -40oC Ambient |
|
Temperature Control |
Normally Set at 40 oC,
controlled with 3 term controller |
Electrical
|
Power Supply: |
100 or 254 VAC, 50 - 60 Hz |
|
Enclosure: |
IP65 Floor mounting integrally insulated GRP enclosure, with
stainless steel fittings. |
|
Electrical Connection: |
Power: Junction Box, Option Auxillary Fuse Box
Signal: Main EExd enclosure
Alarms: Main EExd enclosure
Laptop: Auxillary Junction Box |
|
Controls: |
Internal: Membrane Keypad, settings stored in EEPROM |
| |
External: Option Push Buttons on EExd Enclosure |
|
|
|
|
Impulse Connections |
Client to specify |
Certification
|
Electrical |
ATEX |
| |
FM |
|
CSA |
|
Material |
EN 10204 3.1b Over 4 BarG |
|
NACE |
Material statement from IMA, User to refer to standards |
|