Channels Single channel
   
Dimensions • Size: 238 x 138 x 38 mm (9.4 x 5.5 x 1.5 in)
  • Weight, electronics: 1.13 kg (2.5 lb)
  • Weight, sample system: 1.8 kg(4 lb)
Enclosure Type 4X, IP67
Power (AC Version) 120 or 230 VAC, 50-60 Hz power supply with appropriate power cord
   
Battery Pack (Option on AC Version) • Requires separate switchable input battery charger: 115 or 230 VAC, 50/60 Hz
  • Battery life: 15 to 24 hours depending on type of probe; battery life reduced when operating below 0°C (32°F)
  • Requires approximately three hours for full battery recharge
   
Memory FLASH memory
   
Operating Temperature –10° to 50°C (14° to 122°F)
  To ensure maximum battery life, it is not recommend storing at temperatures exceeding 35°C(95°F ) for longer than one month.
   
Keypad 25-key, rubberized, tactile membrane
   
Display 240 x 200 pixel, graphic backlit LCD display
   
Printer/Terminal Output Infrared communication port
   
Electronics  
Cables Cable type dependent on probe type: M Series, TF Series, or Moisture Image Series. LEMO®-to-bayonet connector
   
Cable Length • Standard: 3 m (10 ft)
  • Optional: Consult GE for other lengths
   
European Compliance Complies with EMC Directive 89/336/EEC, EN61326 Annex A
  For EN61000-4-3, the PM880 meets performance Criteria A and in a limited number of frequencies, performance Criteria B per EN61326
   
Site Parameter Programming Menu-driven, graphic, operator interface uses keypad and soft-function keys. Online help functions. Memory storage for saving site parameters.
   
Data Logging Memory capacity to log over 100,000 moisture data points. Programmable keypad for log units, update times, and start and stop times.
   
Display Functions Displays measurements and logged data in alphanumeric or graphic format. Language options: Dutch, English, French, German, Italian, Portuguese,
Russian, Spanish
   
Display Units • Moisture: DP temperature, ppmv, ppmw, % RH, lb/MMSCF, and others
  • Temperature: °C, °F, and °K
  • Pressure: bar, psig, kPa (gauge), kg/cm2 (gauge), and others
   
Operational Compatibility Compatible with all GE aluminum oxide moisture probes: M Series, TF Series, and Moisture Image Series. Each probe type requires a different type of I/O cable.
   
Calibration Moisture sensors are computer-calibrated to National Institute of Standards and Technology (NIST) traceable moisture concentrations.
   
Overall Calibration Range Capability –110°C to 60°C (-166°F to 140°F)
   
Calibration Range Options  • Standard: 20° to –80°C (68°F to -112°F) with data to –110°C (-166°F)
  • Ultralow: –50°C to –110°C (-58°F to -166°F)
  • Extended high: 60°C to –80°C (140°F to -112°F) with data to –110°C (-166°F)
   
Accuracy • ±2°C(±1°F )  from 60° to –65°C (140°F to -85°F)
  • ±3°C (±2°F) from –65°C to –110°C (-85°F to -166°F)
   
Repeatability • ±0.5°C (±0.3°F) from 60°C to –65°C (140°F to -85°F)
  • ±1°C (±0.5°F) from –65°C to –110°C (-85°F to -166°F)
   
Operating Pressure 5 μ of Hg to 345 bar (5000 psig) limited by pressure sensor—see pressure measurement specifications
   
Dew/Frost Point Temperature Optional thermistor available for all GE moisture probes
   
Range –30° to 70°C (–22° to 158°F)
   
Accuracy ±0.5°C (±0.3°F) at –30°C (-22°F)
   
Temperature Measurement Optional pressure sensor available for TF Series and Moisture Image Series moisture probes
   
Pressure Measurement  
Ranges • 3 to 21 bar (30 to 300 psig)
  • 4 to 35 bar (50 to 500 psig)
  • 7 to 69 bar (100 to 1000 psig)
  • 21 to 207 bar (300 to 3000 psig)
  • 35 to 345 bar (500 to 5000 psig)
   
Accuracy ±1% at full scale
   
Proof Pressure Three times span of available range up to maximum 518 bar (7500 psig)
   
Sample System
 
The SS880A standard sample system comprises an inlet needle valve, a built-in coalescing filter and sample cell with a bypass needle valve and venting tube, a pressure gauge (various ranges), and an outlet needle valve with venting tube.
   
Wetted Parts 316 SS
   
Operating Pressure Configurations available for 21, 35, 69, 207, or 345 bar(300; 500; 1000; 2000; or 3000 psig ), dependent on pressure gauge)
   
Maximum Pressure Rating • Standard: 207 bar (3000 psig)
  • Optional: 345 bar (5000 psig)
   
Other Options • Inlet pressure regulator, 1 to 345 bar (0 to 500 psig) outlet
  • Armoured flowmeter, 1.3 to 13 SCFH
  • 3-m (10-ft), flexible, PTFE lined, braided–stainless steel hose with 1/4-inch tube connections. Not recommended for dew point temperatures below –75°C (–103°F).

 