| Channels |
Single channel |
| |
|
| Dimensions |
• Size: 238 x 138 x 38 mm
(9.4 x 5.5 x 1.5 in) |
| |
• Weight, electronics:
1.13 kg (2.5 lb) |
| |
• Weight, sample system:
1.8 kg(4 lb) |
|
|
| Enclosure |
Type 4X, IP67 |
| Power (AC Version) |
120 or 230 VAC, 50-60 Hz
power supply with appropriate power cord |
| |
|
| Battery Pack (Option on AC
Version) |
• Requires separate
switchable input battery charger: 115 or 230 VAC, 50/60 Hz |
| |
• Battery life: 15 to 24
hours depending on type of probe; battery life reduced when operating below
0°C (32°F) |
| |
• Requires approximately
three hours for full battery recharge |
| |
|
| Memory |
FLASH memory |
| |
|
| Operating Temperature |
–10° to 50°C (14° to
122°F) |
| |
To ensure maximum battery
life, it is not recommend storing at temperatures exceeding 35°C(95°F ) for
longer than one month. |
| |
|
| Keypad |
25-key, rubberized,
tactile membrane |
| |
|
| Display |
240 x 200 pixel, graphic
backlit LCD display |
| |
|
| Printer/Terminal Output |
Infrared communication
port |
| |
|
| Electronics |
|
| Cables |
Cable type dependent on
probe type: M Series, TF Series, or Moisture Image Series. LEMO®-to-bayonet
connector |
| |
|
| Cable Length |
• Standard: 3 m (10 ft) |
| |
• Optional: Consult GE
for other lengths |
| |
|
| European Compliance |
Complies with EMC
Directive 89/336/EEC, EN61326 Annex A |
| |
For EN61000-4-3, the
PM880 meets performance Criteria A and in a limited number of frequencies,
performance Criteria B per EN61326 |
| |
|
| Site Parameter Programming |
Menu-driven, graphic,
operator interface uses keypad and soft-function keys. Online help
functions. Memory storage for saving site parameters. |
| |
|
| Data Logging |
Memory capacity to log
over 100,000 moisture data points. Programmable keypad for log units, update
times, and start and stop times. |
| |
|
| Display Functions |
Displays measurements and
logged data in alphanumeric or graphic format. Language options: Dutch,
English, French, German, Italian, Portuguese,
Russian, Spanish |
| |
|
| Display Units |
• Moisture: DP
temperature, ppmv, ppmw, % RH, lb/MMSCF, and others |
| |
• Temperature: °C, °F,
and °K |
| |
• Pressure: bar, psig,
kPa (gauge), kg/cm2 (gauge), and others |
| |
|
| Operational Compatibility |
Compatible with all GE
aluminum oxide moisture probes: M Series, TF Series, and Moisture Image
Series. Each probe type requires a different type of I/O cable. |
| |
|
| Calibration |
Moisture sensors are
computer-calibrated to National Institute of Standards and Technology (NIST)
traceable moisture concentrations. |
| |
|
| Overall Calibration Range
Capability |
–110°C to 60°C (-166°F to
140°F) |
| |
|
| Calibration Range Options |
• Standard: 20° to –80°C
(68°F to -112°F) with data to –110°C (-166°F) |
| |
• Ultralow: –50°C to
–110°C (-58°F to -166°F) |
| |
• Extended high: 60°C to
–80°C (140°F to -112°F) with data to –110°C (-166°F) |
| |
|
| Accuracy |
• ±2°C(±1°F ) from
60° to –65°C (140°F to -85°F) |
| |
• ±3°C (±2°F) from –65°C
to –110°C (-85°F to -166°F) |
| |
|
| Repeatability |
• ±0.5°C (±0.3°F) from
60°C to –65°C (140°F to -85°F) |
| |
• ±1°C (±0.5°F) from
–65°C to –110°C (-85°F to -166°F) |
| |
|
| Operating Pressure |
5 μ of Hg to 345 bar
(5000 psig) limited by pressure sensor—see pressure measurement
specifications |
| |
|
| Dew/Frost Point Temperature |
Optional thermistor
available for all GE moisture probes |
| |
|
| Range |
–30° to 70°C (–22° to
158°F) |
| |
|
| Accuracy |
±0.5°C (±0.3°F) at –30°C
(-22°F) |
| |
|
| Temperature Measurement |
Optional pressure sensor
available for TF Series and Moisture Image Series moisture probes |
| |
|
| Pressure Measurement |
|
| Ranges |
• 3 to 21 bar (30 to 300
psig) |
| |
• 4 to 35 bar (50 to 500
psig) |
| |
• 7 to 69 bar (100 to
1000 psig) |
| |
• 21 to 207 bar (300 to
3000 psig) |
| |
• 35 to 345 bar (500 to
5000 psig) |
| |
|
| Accuracy |
±1% at full scale
|
| |
|
| Proof Pressure |
Three times span of
available range up to maximum 518 bar (7500 psig) |
| |
|
Sample System
|
The SS880A standard
sample system comprises an inlet needle valve, a built-in coalescing filter
and sample cell with a bypass needle valve and venting tube, a pressure
gauge (various ranges), and an outlet needle valve with venting tube. |
| |
|
| Wetted Parts |
316 SS |
| |
|
| Operating Pressure |
Configurations available
for 21, 35, 69, 207, or 345 bar(300; 500; 1000; 2000; or 3000 psig ),
dependent on pressure gauge) |
| |
|
| Maximum Pressure Rating |
• Standard: 207 bar (3000
psig) |
| |
• Optional: 345 bar (5000
psig) |
| |
|
| Other Options |
• Inlet pressure
regulator, 1 to 345 bar (0 to 500 psig) outlet |
| |
• Armoured flowmeter, 1.3
to 13 SCFH |
| |
• 3-m (10-ft), flexible,
PTFE lined, braided–stainless steel hose with 1/4-inch tube connections. Not
recommended for dew point temperatures below –75°C (–103°F). |