Home

|| Home ||


DewLogic View Dewpoint Transmitter

Technical Specifications

Moisture Range

–110 to 20°C (Standard)

 

Options
–110 to –50°C
–80 to 20°C
–80 to –30°C
–30 to 20°C 
–60 to 40°C
–150 to 70°F
–150 to –40°F
–40 to 70°F
–100 to 0°F
–50 to 50°F
0 to 100°F
0 to 10 PPMv
0 to 100 PPMv
0 to 1000 PPMv
0 to 10,000 PPMv Note: PPMv ranges based on constant pressure, provided at time of order placement.

Operating Temperature

–40° to 60°C (–40° to 140°F)

Storage Temperature

70°C (158oF) maximum. The probe should be stored with the plastic cover and desiccant packet threaded onto the probe. Store in a cool, dry environment.

Warm-up Time

Meets specified accuracy in 3 minutes

Accuracy

+/- 2°C from –65o to 40°C (-85o to 104oF)
+/- 3°C from –80o to –66°C (-112o to -87oF)

Repeatability

+/- 0.5°C from –65o to 40°C (-85o to 104oF) dew/frost point
+/- 1.0°C from –80o to –66°C (-112o to -87oF)

Response Time

Less than 5 seconds for 63% of a step change in moisture content in either wet-up or dry-down cycle

   

Electronics

 

Supply voltage

7 to 28 VDC (loop-powered, customer supplied)

Output

4 to 20 mA

Output Resolution

0.01 mA Max.

 

Loop R = 50 Ohms x (PSV–7), where PSV = Power Supply Voltage
Example: Given a 24 VDC Power Supply, Max. Loop R = 50 Ohms x (24–7) = 850 Ohms

Cable

2 m, standard (consult IMA for custom lengths)

   

Mechanical

 

Sample Connection

3/4-16 straight male thread with o-ring

Operating Pressure

5 Micron Hg to 5,000 psig (345 bar)

Enclosure

NEMA 4X/IP67

European Compliance

Complies with the following: • EMC Directive 89/336/EEC and PED 97/23/EC for DN<25 • EN 61326:1998 Class A, Annex A, Continuous Unmonitored Operation (For EN 61000-4-3 transmitter meets performance criteria A and in a number of frequencies, criteria B per EN 61326)

Dimensions

Overall: 6.76 x 1.13 in (17.17x 2.87 cm) Electronics with cable: 4.08 x 1.13 in. diameter (10.36 x 2.87 cm)

Weight

140 grams (5 oz)

   

Moisture Sensor

 

Sensor Type

Thin-film aluminum oxide moisture sensor probe

Calibration

Each sensor is individually computer-calibrated against known moisture concentrations, traceable to NIST

Calibration Interval

Sensor recalibration recommended every 6 to 12 months depending on application

Calibration Data

Factory-calibrated, stored in FLASH

Flow Rate

Static to 10,000-cm/s linear velocity at a pressure of 1 atm

   
 

All specifications are subject to change for product improvement without notice.

   

Warranty:

1 Year