Moisture Range |
–110 to 20°C (Standard) |
Options |
|
Operating Temperature |
–40° to 60°C (–40° to 140°F) |
Storage Temperature |
70°C (158oF) maximum. The probe should be stored with the plastic cover and desiccant packet threaded onto the probe. Store in a cool, dry environment. |
Warm-up Time |
Meets specified accuracy in 3 minutes |
Accuracy |
+/- 2°C from –65o to 40°C (-85o to 104oF)
|
Repeatability |
+/- 0.5°C from –65o to 40°C (-85o to 104oF) dew/frost point |
Response Time |
Less than 5 seconds for 63% of a step change in moisture content in either wet-up or dry-down cycle |
Electronics |
|
Supply voltage |
7 to 28 VDC (loop-powered, customer supplied) |
Output |
4 to 20 mA |
Output Resolution |
0.01 mA Max. |
Loop R = 50 Ohms x (PSV–7),
where PSV = Power Supply Voltage
|
|
Cable |
2 m, standard (consult IMA for custom lengths) |
Mechanical |
|
Sample Connection |
3/4-16 straight male thread with o-ring |
Operating Pressure |
5 Micron Hg to 5,000 psig (345 bar) |
Enclosure |
NEMA 4X/IP67 |
European Compliance |
Complies with the following: • EMC Directive 89/336/EEC and PED 97/23/EC for DN<25 • EN 61326:1998 Class A, Annex A, Continuous Unmonitored Operation (For EN 61000-4-3 transmitter meets performance criteria A and in a number of frequencies, criteria B per EN 61326) |
Dimensions |
Overall: 6.76 x 1.13 in (17.17x 2.87 cm) Electronics with cable: 4.08 x 1.13 in. diameter (10.36 x 2.87 cm) |
Weight |
140 grams (5 oz) |
Moisture Sensor |
|
Sensor Type |
Thin-film aluminum oxide moisture sensor probe |
Calibration |
Each sensor is individually computer-calibrated against known moisture concentrations, traceable to NIST |
Calibration Interval |
Sensor recalibration recommended every 6 to 12 months depending on application |
Calibration Data |
Factory-calibrated, stored in FLASH |
Flow Rate |
Static to 10,000-cm/s linear velocity at a pressure of 1 atm |
All specifications are subject to change for product improvement without notice. |
|
Warranty: |
1 Year |