When accurate measurements are required where entrained
liquids, particulates, or varying conditions are present a sample system is often
required. Particulates pose an obvious problem - erosion of the sensor leading to
inevitable drift and premature failure - hence filtration is generally required.
Liquid entrainment has less obvious effects; most substances are either hydrophilic or
hydrophobic (water attracting or repelling). Should these coat the sensor the balance with
the measured gas will alter - leading to an apparent offset. To counter this problem
coalescing filters are used to remove the liquids, and additional heat will ensure the
remainder will not condense out.
With analysers sometimes being located some distance from the tapping point slow reaction
times, or the effects of Transpiration can
cause problems. To reduce the lag or dilute its effects fast by-pass loops are installed.